4. When the exposure warning stops and the program end tone is played, the program
is complete.
5. The calibration image appears on the GUI.
NOTICE! If the calibration fails, the height or tilt of the CEPH primary
collimator needs to be adjusted. See chapter
adjustment
6. Acknowledge the calibration result by pressing the OK button.
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
on page 175 for more information.
OP 3D
CEPH primary collimator
9 Installation