KaVo ORTHOPANTOMOGRAPH OP 3D User And Installation Manual page 159

Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

DEVICE SETTINGS:
Friendly name
3D Metal Artefact
Reduction
(3D devices only)
Force device language
Default to Scout
before 3D
(3D devices only)
DAP corrections
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
Set a name for the device, which is shown next to the
device settings icon and on the imaging software.
Enable or disable MAR, Metal Artefact Reduction. MAR
is used to reduce the effect of metals and other dense
radiopaque objects on the 3D image.
It is recommended to keep MAR activated.
Change the language of the device GUI if needed.
It is recommended not to force the language of the GUI
but change the language profile of the workstation.
Enable or disable automatic Scout image activation when
a 3D imaging modality is selected. It is recommended to
have the Scout image taking enabled.
Change the DAP correction factor, which is used to
calculate the shown DAP value on the GUI. This setting is
preset at the factory and should not be altered in normal
conditions.
There are separate DAP correction options for PAN, 3D
and CEPH imaging.
OP 3D
NOTICE! If the selection is set to "disabled", a
Scout image mode can still be activated manually
in the 3D imaging program selection view.
NOTICE! Changing the DAP correction factor has
no impact on the device radiation output, it is used
for DAP display calibration purposes only.
9 Installation
159
D519207, 19