INSTALLATION SETTINGS:
License
Select country
Door Switch
Demo mode
PAN QC spinal
compensation
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
If the user has purchased a licensable upgrade for
optional imaging programs, for example 9x14 3D FOV,
they can be activated through this field.
Type the provided license key to the field and press Save
to activate the option.
Select the installation country of the device. The selection
configures the device according to country specific
regulations, for example change the maximum selectable
mA for imaging.
If an optional Door Switch has been installed to the
device according to chapter
page 86, it has to be activated here.
Used for setting the device into demonstration mode
where the radiation production of the device is disabled
while maintaining all other functionalities.
Enable spinal compensation for PAN QC program.
OP 3D
NOTICE! For PAN QC to be used with 1.8mm
Cu filter and CEPH QC to be used with 0.8mm Cu
filter, select Germany.
Back panel connectors
NOTICE! Demo mode is intended for exhibition
use only.
9 Installation
on
161
D519207, 19