10 Technical data
Table 2: Electromagnetic immunity IEC 60601-1-2 Ed4
PCX-1 is suitable for use in the specified electromagnetic environment. The purchaser or user of
PCX-1 should assure that it is used in an electromagnetic environment as described below:
Immunity Test
Electrostatic
discharge (ESD)
IEC61000-4-2
Electrical fast
transient/burst IEC
61000-4-4
Surge
IEC61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC61000-4-11
Power frequency
(50/60Hz) magnetic
field IEC 61000-4-8
NOTICE! U
214
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
IEC60601-1-2
Compliance Level
Test Level
± 8 kV for contact
± 8 kV for contact
discharge
discharge
±2, 4, 8, 15 kV for
±2, 4, 8,15 kV for
air discharge
air discharge
±2 kV for power
±2 kV for power
supply lines (100
supply lines
kHz)
±1 kV for input/
±1 kV for input/
output lines
output lines
(100kHz)
± 0.5, 1 kV
± 0.5, 1 kV
differential mode
differential mode
± 0.5, 1, 2 kV
± 0.5, 1, 2 kV
common mode
common mode
-0 % U mains; 0,5
-0 % U mains; 0,5
cycle At 0°, 45°,
cycle At 0°, 45°,
90°, 135°, 180°,
90°, 135°, 180°,
225°, 270° and
225°, 270° and
315°
315°
-0 % U mains; 1
-0 % U mains; 1
cycle, At 0°
cycle, At 0°
-70 % U mains;
-70 % U mains;
25/30 cycles At 0°
25/30 cycles At 0°
-0 % U mains;
-0 % U mains;
250/300 cycle At 0°
250/300 cycle At 0°
30A/m
30 A/m
is the a.c. mains voltage prior to application of the test level.
T
Electromagnetic Environment
Floors are wood, concrete, or ceramic
tile, or floors are covered with synthetic
material and the relative humidity is at
least 30 percent.
Mains power quality is that of a typical
commercial and/or hospital environment
Mains power quality is that of a typical
commercial and/or hospital environment.
Mains power quality is that of a
typical commercial and/or hospital
environment. If the user of PCX-1
requires continued operation during power
mains interruptions, it is recommended
that PCX-1 be powered from an
uninterruptible power supply.
Power frequency magnetic fields are at
levels characteristic of a typical location
in a typical commercial and/ or hospital
environment.
The device shall not be used closer than
15cm to sources of 50/60Hz magnetic
field.
ORTHOPANTOMOGRAPH
OP 3D
D519207, 19