Ceph Generator Calibration; Ceph Tube Sweep Calibrations - KaVo ORTHOPANTOMOGRAPH OP 3D User And Installation Manual

Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

9 Installation
10. Once the CEPH arm is tilted correctly and aligned with the FH light, tighten the 4
CEPH arm fixing bolts.
11. Finish the calibration by pressing the OK button.
12. Attach the PAN/3D tubehead front cover.

9.6.16 CEPH generator calibration

This program calibrates the CEPH tubehead and generator.
No calibration tools are required when performing this calibration.
NOTICE! This calibration produces X-rays. Protect yourself from radiation.
1. Select CEPH generator calibration from the Calibrations menu.
2. Protect yourself from radiation.
3. Press and hold the exposure button down to start the calibration.
4. When the exposure warning stops and the program end tone is played, the program
is complete.

9.6.17 CEPH tube sweep calibrations

These two programs are used to calibrate the CEPH tubehead rotation and the CEPH
primary collimator.
No calibration tools are required when performing these calibrations.
NOTICE! This calibration produces X-rays. Protect yourself from radiation.
Perform the calibrations in this order, the instructions are the same for both programs:
• CEPH tube sweep calibration
• CEPH tube sweep calibration 2
1. Select CEPH tube sweep calibration from the Calibrations menu.
2. Protect yourself from radiation.
3. Press and hold the exposure button down to take the calibration image.
134
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
ORTHOPANTOMOGRAPH
D519207, 19
OP 3D

Advertisement

Table of Contents
loading

Table of Contents