9 Installation
3. To move the collimator in UP/DOWN direction, turn the collimator height
adjustment screw:
a) Clockwise to move the collimator downwards.
b) Counter clockwise to move the collimator upwards.
NOTICE! The GUI informs you how much the collimator needs to be adjusted
and in which direction. One full turn of the screw (360°) equals about 0.7 mm.
4. To rotate the collimator CLOCKWISE (CW)/COUNTER CLOCKWISE (CCW),
loosen the 2 upper locking screws (1), the lower locking screw (2) and turn the
adjustment screw (3):
a) Clockwise to turn the collimator counter clockwise.
b) Counter clockwise to turn the collimator clockwise.
NOTICE! You may also loosen the screws 2 and 3, turn the collimator to the
required position by hand and then tighten the screws.
NOTICE! Use the angle scale on the tubehead to determine how much you
have rotated the collimator.
5. After the adjustments, tighten the locking screws and redo the collimator calibration
to verify the adjustment.
6. Re-install the detached covers.