9 Installation
5. Turn the head support to PA position.
6. Pull the left hand side of the upper head support cover over the head support mount.
7. Turn the head support to LAT position and remove the cover.
8. Loosen the ear rod tilt locking screws (2 pcs) and adjust the eccentric screw to
adjust the ear holder tilt as required.
NOTICE! The ear rod assembly has a tilt amount indicator to be used as a
guide.
9. Verify the correct ear rod tilt by performing the CEPH position calibration.
10. Re-install the detached covers.
NOTICE! Remember to press the lower head support cover back against the
upper cover to re-attach it.