9 Installation

9.5.2 Standalone Configuration

Standalone Configuration
In standalone configuration, the unit is connected directly to a workstation and the unit
then reserves the whole network interface card. The unit and the workstation function
as a single standalone system with no access to other networks. If it is necessary to be
connected to another network or to a server, the workstation will require two network
interface cards.
In standalone configuration, both the ETHERNET interface of the workstation and
the device IP address need to be set to have IP addresses from the same subnet, for
example 192.168.1.1 for the PC and 192.168.1.2 for the device. The recommended
Subnet mask is 255.255.255.0.
A static IP address can be configured to the device through the device driver. See
software documentation for more information on how to configure the imaging software
for device connection required for the static IP address configuration.
NOTICE! When configuring a static IP address it must be set for both the device
and the PC. The device may have initial local host IP Address (169.xxx.xxx.xxx)
but this is not static unless set. It is not recommended to use addresses from the
local host address range (169.xxx.xxx.xxx) as static IP addresses.