7 Maintenance
9. Workstation: Visually evaluate and inspect the 3D image for visual defects such as
artifacts using the 3D imaging software.
NOTICE! The device determines if the QC image is PASSED or FAILED
according to measured data, not based on the visible image quality.
NOTICE! You should also compare the new QC image to the reference image
taken during the installation or the latest service. Doing this helps you ensure
that the image quality has remained constant.
10. If the image fails or any visual defects are apparent, redo the QC program. If it
fails again, recalibrate the device according to
contact service.
11. Detach the 3D QC phantom and the 3D QC phantom holder from the device.
76
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
Calibration procedure
on page 69 or
ORTHOPANTOMOGRAPH
™
OP 3D
D519207, 19