To rotate the collimator LEFT/RIGHT:
1. Loosen the indicated collimator assembly installation screws (3 pcs).
2. Rotate the collimator assembly by hand. The amount of rotation can be inspected
from the scale on the tubehead.
4. After the adjustments, re-tighten the loosened screws and rerun the CEPH Tube
Sweep Calibration to verify the adjustment.
5. Re-install the detached covers.