KaVo ORTHOPANTOMOGRAPH OP 3D User And Installation Manual page 177

Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

To rotate the collimator LEFT/RIGHT:
1. Loosen the indicated collimator assembly installation screws (3 pcs).
2. Rotate the collimator assembly by hand. The amount of rotation can be inspected
from the scale on the tubehead.
4. After the adjustments, re-tighten the loosened screws and rerun the CEPH Tube
Sweep Calibration to verify the adjustment.
5. Re-install the detached covers.
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
OP 3D
9 Installation
177
D519207, 19

Advertisement

Table of Contents
loading

Table of Contents