4. Select PAN exposure tests, CEPH exposure tests or 3D exposure tests tab.
The PAN exposure test and CEPH exposure test programs allow adjustment of
the exposure values kV and mA for the tests.
The 3D exposure test program allows selection of different radiaton modes High/
ENDO Resolution (Constant), Standard Resolution (Pulse 10ms) or LDT resolution
(Pulse 5ms) and adjusting the related mA.
NOTICE! When using the PAN exposure test program, the X-Ray beam is
adjusted wider (over 2 cm) than in normal imaging programs to allow easier
positioning of the measurement device.
5. Attach a proper measurement device to the sensor so that the radiation beam hits
the detector of the measurement device.
6. Press and hold the exposure button down for the duration of the test.

9.10.5 Remote exposure switch installation

A wall-mountable remote exposure switch, with integrated exposure warning light, is
available as an option for the device.
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
OP 3D
9 Installation
167
D519207, 19