9 Installation
9.8.3 CEPH QC (optional)
NOTICE! This program is available only for devices with cephalometric (CEPH)
modality.
NOTICE! The CEPH QC is an optional, yet recommendable, procedure and is
mandatory to be performed where local regulations require it.
NOTICE! The device's country selection needs to be configured per local
requirements before performing the QC programs. See chapter
page 158.
1. GUI: Select CEPH QC program from the Quality Control menu.
2. Turn the CEPH head support to PA position.
3. Turn the nasion support aside.
4. Attach the CEPH QC phantom holder to the ear rods so that the indicator on the
holder faces the CEPH tubehead.
5. Attach the 2D QC test phantom to the CEPH QC phantom holder.
152