7 Maintenance
3. Press and hold the exposure button down to take the calibration image.
4. When the exposure warning stops and the program end tone is played, the program
is complete.
5. The calibration image appears on the GUI.
6. Acknowledge the calibration result by pressing the OK button.
7.2.3.3 3D geometry calibrations
These programs create the calibration data for reconstructing 3D images.
NOTICE! This calibration produces X-rays. Protect yourself from radiation.
Perform the calibrations in this order, the instructions are the same for all programs:
• 3D geometry calibration, 5x5 FOV
• 3D geometry calibration, 6x9 FOV
• 3D geometry calibration, 9x11 FOV
• 3D geometry calibration, 9x14 FOV
Geometry Calibration Phantom is required when performing these calibrations.
1. Attach the Geometry Calibration Phantom to the device.
2. Select a 3D geometry calibration from the Calibrations menu.
3. Protect yourself from radiation.
70