Advertisement

6 Using the device
4. Place disposable covers on the patient positioning accessories.
5. Adjust the device's height to approximately match the patient's height by sliding a
finger on the carriage up/down slider.
6. Open the head support locking lever on the lower shelf, push the head support
towards the mirror and lock it leaning forward.
7. Tell the patient to remove their glasses, hearing aids, removable dentures, jewellery,
hair clips and all other things that can cause artifacts to the image.
8. Protect the patient from radiation according to the local regulations, for example with
a lead apron and a thyroid shield.
9. Guide the patient to the device and instruct to stand as straight and as tall as
possible.
10. Ask the patient to grab the patient handles, place their chin on the Chin Rest and to
press their lip against the Lip Support.
NOTICE! If the patient has wide shoulders that could collide with the rotating
unit, ask the patient to cross their arms when holding the handles in order to
contract the soulders.
11. Ask the patient to take a step forward to straighten their spine.