9.2.3 Fixing hardware and tools

The following tools and hardware are required to install and set up the device. These are
not included in the delivery, unless otherwise stated.
Fixing hardware
NOTICE! Fixing hardware for floor and wall fixing are not included in the delivery.
The type and length of the fixing hardware to be used to fix the device to the wall and
floor depend on the wall and floor materials. The fixing hardware must be selected so
that they are of correct type for the wall and floor materials and they withstand at least
1500 N pull-out force.
NOTICE! If you wish to drill the floor fixing holes through the column floor plate
fixing holes, use max. 10 mm/0.4 in diameter floor fixing hardware.
WARNING! Ensure that the fixing hardware, the wall and the floor the device is
fixed to, can withstand at least 1500 N pull-out force.
Installation tools
• Electric drill
• Spanners size 13 & 17 mm
• Allen keys 3, 5 & 6 mm
• Flat head screwdriver
• Spirit level
• Pliers and wire cutter
• Scissors/knife
• Torque wrench
ORTHOPANTOMOGRAPH