6.4.2 Patient positioning for Cephalometric imaging

NOTICE! The device can be used to take images of both standing and seated
patients. It is recommended to have very tall patients seated for easier
positioning.
1. Press HOME button on the patient positioning panel. The device moves to CEPH
HOME (Patient-In) position.
NOTICE! Ensure that the device has been set up correctly for the required
examination.
2. The device goes to ready state and the status indicator light turns green.
NOTICE! If the status indicator light does not turn green, look for more
information on the GUI.
3. Turn the CEPH head support to Lateral or PA position, according to the required
examination.
4. To ease the patient entry you may turn the nasion support an the ear rods aside and
pull them to their outmost position.
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
Lateral projection
OP 3D
Posterior-Anterior (PA) projection
6 Using the device
53
D519207, 19