9 Installation
22. Connect the cables inside the carriage.
3-pin connector
NOTICE! The tubehead cable set connectors can be installed only to their
correct locations.
NOTICE! The CEPH screw kit contains the screw and star washer for the
grounding cable.
23. Attach the tubehead cable set cables to grounding points with clamps:
• 3-pin connector cable with a size 4 clamp (upper) and a size 5 clamp (lower).
• 14-pin connector cables with size 3 (thicker cable) and size 2 (thinner cable)
clamps. Both clamps can be connected to the same grounding point.
24. Re-attach the mirror base plate, tighten the previously loosened screws (3 pcs) and
re-attach the mirror opto cable.
25. Guide the remaining gray cable from the CEPH tubehead through the opening on the
CEPH arm and connect it to the lower connector on the rear of the device.