7 Maintenance
5. The calibration image appears on the GUI.
6. Acknowledge the calibration result by pressing the OK button.
7.2.3.5 CEPH pixel calibration
This program calibrates the sensor for Cephalometric imaging.
No calibration tools are required when performing this calibration.
NOTICE! This calibration produces X-rays. Protect yourself from radiation.
1. Turn the CEPH head support to LAT position.
2. Move the nasion support to its outmost position an turn it down.
3. Select CEPH pixel calibration from the Calibrations menu.
4. Protect yourself from radiation.
5. Press and hold the exposure button down to take the calibration image.
6. When the exposure warning stops and the program end tone is played, the program
is complete.
7. The calibration image appears on the GUI.
8. Acknowledge the calibration result by pressing the OK button.
72
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
ORTHOPANTOMOGRAPH
D519207, 19
OP 3D