7. The QC image appears on the workstation, showing the result of the check.
8. Workstation: Visually evaluate and inspect the 3D image for visual defects such as
artifacts using the 3D imaging software.
NOTICE! The device determines if the QC image is PASSED or FAILED
according to measured data, not based on the visible image quality.
9. If the image fails or any visual defects are apparent, redo the QC program. If it fails
again, recalibrate the device according to
contact service.
10. Save the passed QC image as a reference image for the user and let the user know
the reference image location.
11. Detach the 3D QC phantom and the 3D QC phantom holder from the device.