4 User Interfaces
IMAGING PROGRAM SETTINGS:
Test mode
Patient size
End Study
STATUS MESSAGE BAR:
Message bar
Device status indicator
20
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
Indicates if Test mode is active.
Press Test mode icon to enable/disable device radiation production.
Test mode can be used for example to demonstrate the device
movements.
You can also use the patient positioning panel to enable/disable the
Test mode.
Indicates the currently selected patient size preset.
Press the Patient size selection icon to open a list of patient sizes;
Small, Medium & Large.
Press on the patient size icon to activate the preset.
Selecting a patient size, according to the scanned patient, adjusts
the kV and mA to preset levels.
NOTICE! Patient size presets can be adjusted from device
settings.
Press End Study icon to finalize the imaging procedure and return to
the main view.
The message bar provides messages and guidance to the user. Press
on the bar to show messages if there are more than the ones shown.
Status message bar shows the device status with a colored Device
status indicator.
• GREEN: Device is ready for imaging.
• GRAY: Device is not ready for imaging, follow the instructions on
the status message bar.
• YELLOW: Device is radiating.
• BLUE: Device is in error state or pending user action. Details are
shown on the status message bar.
ORTHOPANTOMOGRAPH
D519207, 19
OP 3D