9 Installation

9.6.15 CEPH head support calibration

This program is used to calibrate the different positions of the CEPH head support and
nasio support. During this calibration, the alignment of the CEPH arm is also checked and
adjusted.
No calibration tools are required when performing this calibration.
1. Remove the PAN/3D tubehead rear cover by detaching the 3 indicated screws and
pull the cover outwards. The screw on the middle of the cover has a screw cover cap
over it.
2. Remove the screws (4 pcs) holding the PAN/3D tubehead front cover in place and
remove the cover.
3. Select CEPH head support calibration from the Calibrations menu on the GUI.
4. Press HOME button on the patient positioning panel. The device moves to CEPH
HOME position.
5. Turn the CEPH head support to LAT and PA positions for a couple of times.
132
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37