6. Move the nasion support inwards and outwards to it's mechanical limits for a couple
of times.
7. Check that the CEPH FH light goes through both of the markers on the 2D sensor
cover. This check is done to verify that the CEPH FH light is correctly aligned.
NOTICE! If the light doesn't hit both of the marks, adjust the FH light
alignment. See chapter
NOTICE! The tubehead front cover must be off the device for the light to hit
the markings.
8. Turn the CEPH head support to LAT position and check that the FH light hits the tip
of the sensor side ear rod.
9. If the FH light doesn't hit the tip of the ear rod, adjust the tilt of the CEPH arm. The
tilt of the CEPH arm can be adjusted with the leveling bolt.
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
CEPH FH light
OP 3D
on page 189 for more information.