4. Cut the protective plastic aside to reveal the carriage base plate, carriage installation
rail and back panel connectors.
5. Check that the device's mains power configuration is applicable to the installation
location.
NOTICE! The device is delivered with two 10A fuses installed. If installing to
other than 230V mains voltage the fuses need to be changed. See chapters
Mains power configuration
for more information.
NOTICE! If mains connection type needs to be changed, see chapter
power configuration
6. Move the carriage against the column.
NOTICE! If the device is installed in an angle (e.g. corner installation) the
plywood supports may come in the way of the installation and need to be
removed at this point. See more details in chapter
203.
NOTICE! If the device is installed in a tight space and the column has been
attached to the floor only with the pivoting screw, you can rotate the column
to a more suitable position for the carriage installation. See chapter
installation
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
on page 163 for more information.
on page 87 for information.
OP 3D
on page 163 and
Changing the fuses
9 Installation
on page 165
Mains
Device dimensions
on page
Column
D519207, 19
97