10 Technical data

10.2 Imaging program specifications

NOTICE! The values given in this chapter are nominal values and apply to
220-240 VAC devices. The technical factors and their ranges are limited and may
differ from the given values when using a device configured to 100 - 120 VAC.
Panoramic programs & technical factors
Magnification factor: 1.3
Program
Standard Panoramic
Segmented Standard
Panoramic
Pediatric Panoramic
Segmented Pediatric
Panoramic
Bitewing
TMJ, Lateral
Cephalometric programs & technical factors
Magnification factor: 1.145
Program
Lateral
projection
Pediatric lateral
projection
PA projection
Carpus view
Scout programs & technical factors
FOV size
5 x 5
6 x 9
9 x 11
9 x 14 (optional)
NOTICE! Tube voltage is fixed to 95 kV
200
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
Field of view height
2D/3D:
143.5 mm / 5.6 in
2D:
148.5 mm / 5.8 in
114 mm / 4.5 in
103 mm / 4 in
2D/3D:
143.5 mm / 5.6 in
2D:
148.5 mm / 5.8 in
Field of
Field of
view width
view height
160-260 mm
223 mm
6.7 - 10.2 in
8.9 in
180 mm
7.1 in
200 mm
223 mm
7.9 in
8.9 in
200 mm
223 mm
7.9 in
8.9 in
Resolution
Pixel size
Scout
198 μm
Scout
198 μm
Scout
198 μm
Scout
198 μm
kV range
mA range
60 - 70 kV
2.0 - 16.0 mA
73 - 81 kV
2.0 - 14.0 mA
85 - 90 kV
2.0 - 12.5 mA
kV range
mA range
77 - 81 kV
4.0 - 14.0 mA
85 - 90 kV
4.0 - 12.5 mA
95 kV
4.0 - 11.0 mA
77 - 81 kV
4.0 - 14.0 mA
85 - 90 kV
4.0 - 12.5 mA
95 kV
4.0 - 11.0 mA
60 - 73 kV
2.0 - 12.5 mA
mA range
4.0 - 12.5 mA
4.0 - 12.5 mA
4.0 - 12.5 mA
4.0 - 12.5 mA
ORTHOPANTOMOGRAPH
Exposure time
9.0 s
1.4-9.0 s
9.0 s
1.4-9.0 s
5.20 s (2.6s + 2.6s)
4.0 s (2.0s + 2.0s)
Exposure time
6.5 - 10.5 s
8.1 s
8.1 s
Exposure time
0.03 - 0.04 s
0.07 s
0.07 s
0.07 s
OP 3D
D519207, 19