9 Installation
10. Save the passed QC image as a reference image for the user and let the user know
the reference image location.
11. Detach the PAN QC phantom holder and 2D QC test phantom from the device.
12. Detach the copper filter from the PAN/3D tubehead.
9.8.2 3D QC
NOTICE! This program is available only for devices with 3D modality.
1. GUI: Select 3D QC program from the Quality Control menu.
2. Attach the 3D QC phantom holder and the 3D QC phantom to the device.
3. Protect yourself from radiation.
4. Press and hold the exposure button down to take the QC image.
5. The QC image preview appears on the GUI.
6. Acknowledge the preview by pressing the OK button on the GUI.
150
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37