5 Operation
5.3 Utility programs
▶ Choose ENTER to store the chosen value.
Choosing "no" will cause you to exit the utility program.
Choosing "yes" will complete the procedure.
▶ Press the back button.
This takes you out of the utility program and resets all settings.
Note
Reset = Resets all values to the original factory setting

5.3.7 "Automatic sensitivity adjustment" utility program (2300)

▶ Using the the up or down button, select the "Automatic sensitivity adjustment"
utility program and choose ENTER.
▶ Using the up or down button, select the required workstation (AP1, AP2 or AP3).
The unit to be measured is either switched off or on standby.
▶ Press ENTER
The message "Motor off" appears on screen.
The lower current level is measured. The message on screen switches to "Motor
on".
▶ Switch on the unit to be measured or start up the handpiece at a low speed.
Once the measurement is complete, the following three values are displayed on the
bottom row: a lower limit, an upper limit and a threshold value calculated from the
two values, which, when reached, switches on the extractor unit.
▶ Press ENTER.
This stores the calculated threshold value.
If you do not want to adopt the threshold value, press the back button.
▶ Press the back button
This takes you out of the utility program.