9. CEPH devices only: Determine the installation angle of the device and configure
the CEPH wall bracket arm length accordingly.
NOTICE! The space requirements for 3 example installation angles are shown
in chapter
NOTICE! A full scale installation template is supplied with the device to help
selecting the optimal installation position.
NOTICE! The length of the wall bracket arm can be adjusted slightly for best
fit.
NOTICE! A column rotation scale is located in the column side wall bracket.
A. DEVICE INSTALLED IN AN ANGLE - CLOSEST TO THE REAR WALL
Use the short wall bracket arm in the shortest position (250 mm). The wall
bracket is supplied in this configuration.
Short wall bracket arm in the shortest position enables column rotation between
25-47°.
This wall bracket assembly enables the device to be installed as close to the rear wall
as possible.
B. DEVICE INSTALLED IN AN ANGLE - CLOSE TO THE REAR WALL WITH
MORE ADJUSTABILITY
Use the short wall bracket arm in the longest position (300 mm).
Short wall bracket arm in the longest position enables column rotation between
20-51°.
To change the wall bracket arm length:
a) Detach the screws (4 pcs) from the sides of the arm.
b) Pull the wall bracket arm outwards to 300 mm mark.
c) Re-install the screws (4 pcs).
This wall bracket assembly enables the CEPH tubehead to be installed as close to the
rear wall as possible.
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
Cephalometric device dimensions
OP 3D
on page 206.