9.2 Pre-installation requirements

Familiarize yourself with this manual and the chapter
page 10 to ensure the safe installation of the device.

9.2.1 Installation location requirements

WARNING! Ensure that each of the wall mount fixing screws and the wall can
withstand pull-out force of at least 1500 N.
• The place where the device is to be installed and the position from where the user
takes images must be correctly shielded from the radiation that is generated when the
device is operated. Follow the local radiation and safety requirements.
• The device must be fixed to the wall and the floor.
NOTICE! If the device cannot be fixed to the floor, install the device to an
exhibition stand. See chapter
instructions.
• The wall material should be suitable for fixing the device. If the wall is made of a
weak material, you may have to use a reinforcing plate on the rear side of the wall to
hold the fixing hardware.
• Make sure that the floor, where the device is to be installed, can support its weight.
To avoid the device from tipping over, fix the device with floor bolts appropriate to the
floor material. The floor bolts and the floor material must withstand pull-out force of
at least 1500 N.
• Do not install the device in environments where corrosive or explosive vapours or
flammable anaesthics are present.
• Special steps regarding EMC need to be taken when installing the device. For more
information, refer to the chapter
213.
• The device is supplied with a 3m (10ft) long power cord. Ensure that the connected
power cord and ehternet cables are long enough, as they need to move along with the
device's up/down movements.
NOTICE! It's recommended to route the power feed and Ethernet connection
to behind the device to ensure unobstructed movement of the cables.
• It's recommended to use a max. 30 A circuit breaker with the device.
• Maximum allowed mains line impedance is 0.2 Ω.
• For permanent installation, a separate lockable mains switch (not supplied) is required
to be installed to the mains feed.
• Recommended mains over-current releases:
100-120V: 16A
220-240V: 10A
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
OP 3D
Warnings and requirements
Exhibition stand installation
Electromagnetic Compatibility (EMC) tables
9 Installation
on
on page 170 for
on page
D519207, 19
83

Advertisement

Table of Contents
loading

Table of Contents