9 Installation
15. Attach the wall bracket arm to the column side bracket using the same screws,
washers (2 pcs) and locking nuts (2 pcs) that were previously detached. Do not
tighten the screws fully yet.
16. Adjust the wall bracket arm to the selected installation angle using the scale on the
column side wall bracket as a guide.
17. Tighten the wall bracket angle locking screws to lock the angle of the wall bracket
arm.
18. Position the column to the installation location so that the wall bracket sits properly
against the wall.
19. Measure the distances to the nearest walls and make sure that the selected
installation angle and wall bracket arm length (CEPH devices only) are appropriate.
Adjust the length and the angle of the wall bracket if needed.
NOTICE! See the space requirements from chapter
page 203.
20. Mark the installation screw locations to the wall using the holes in the wall bracket as
a template.
NOTICE! Mark the screw locations to the center of the holes in the wall
bracket to allow some adjustment sideways later on.
NOTICE! The top surface of the wall side bracket should be approximately
790 mm (31.1") from the floor when the wall bracket is horizontal.
NOTICE! If the device cannot be bolted to the floor, an exhibition stand is
required. See chapter
instructions.
92
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
PAN/3D wall bracket
PAN/3D wall bracket
Exhibition stand installation
CEPH wall bracket
CEPH wall bracket
Device dimensions
on page 170 for more
ORTHOPANTOMOGRAPH
on
OP 3D
D519207, 19