9 Installation
NOTICE! The line pair resolution depends also on the other factors than the
device itself, for example the imaging Software configurations. According to the
standards, the distinguishable line pair resolution must be 2.5 LP/mm or better.
12. If the image fails on any of the previously listed criteria, redo the QC program. If it
fails again, recalibrate the device according to
117 or contact service.
13. Save the passed QC image as a reference image for the user and let the user know
the reference image location.
14. Detach the CEPH QC phantom holder and 2D QC test phantom from the device.
15. Detach the copper filter from the CEPH tubehead if it was used.