7.3.2 3D QC
NOTICE! This program is available only for devices with 3D modality.
1. GUI: Go to device settings.
2. GUI: Select 3D QC program from the Quality Control menu.
3. Attach the 3D QC phantom holder and the 3D QC phantom to the device.
4. Protect yourself from radiation.
5. Press and hold the exposure button down to take the QC image.
6. The QC image preview appears on the GUI.
7. Acknowledge the preview by pressing the OK button on the GUI.
8. The QC image appears on the workstation, showing the result of the check.
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
OP 3D
7 Maintenance
75