9 Installation
C. DEVICE INSTALLED PERPENDICULAR TO THE WALL
Use the long wall bracket arm supplied separately in the CEPH installation
accessories box.
Long wall bracket arm enables column rotation between 0-51°.
NOTICE! The column can be rotated in range 0-51° with the long arm, but
installing the column in angles more than 20° with the long arm requires more
space than when using the short arm.
To change the wall bracket arm length:
a) Detach the screws (4 pcs) from the sides of the arm.
b) Replace the short arm with the long arm.
c) Re-install the screws (4 pcs).
This wall bracket assembly enables the CEPH device to be installed in narrower
spaces.
10. Detach the preinstalled installation screws, washers and locking nuts from the
column side bracket.
NOTICE! The large washers are used as transportation support only and can
be disposed of.
11. Remove the wall bracket arm pivot screw and the screw cap by pushing the screw
upwards from the bottom of the wall bracket adjustment block.
90
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
ORTHOPANTOMOGRAPH