9.11.5.3 2D Sensor tilt adjustment
1. Detach the 2D sensor covers according to chapter
page 182.
2. Loosen the 2 sensor position locking screws.
3. Turn the upper sensor position adjustment screw to adjust the sensor tilt. You can
see the amount of tilt adjusted from the alignment holes.
4. Re-tighten the position locking screws after the adjustment.
5. Redo the failed calibration to verify the adjustment.
6. Re-attach the 2D sensor covers.
9.11.5.4 2D Sensor position adjustment
NOTICE! Do not adjust the 2D sensor position for other than PAN only devices.
1. Detach the 2D sensor covers according to chapter
page 182.
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
OP 3D
Removing the 2D sensor covers
Removing the 2D sensor covers
9 Installation