9.11.3 CEPH secondary collimator adjustment

If the CEPH Secondary Collimator Calibration has failed due to the collimator being too
high/low or tilted, adjustment is required.
NOTICE! This adjustment is required only in special cases. After the adjustment,
all cephalometric calibrations related to radiation must be performed again.
1. Remove the PAN/3D tubehead rear cover by detaching the 3 indicated screws and
pull the cover outwards. The screw on the middle of the cover has a screw cover cap
over it.
To move the collimator in UP/DOWN direction:
a. Loosen the collimator plate locking screws.