7 Maintenance
3. A list of available device calibrations is shown with the status of the calibration.
NOTICE! The calibration programs listed on the GUI depend on the device
configuration.
Calibration status indications:
Calibration not performed
or failed.
4. Select a calibration program and follow the instructions on the GUI.
NOTICE! Performed calibrations are indicated with green colour and the
completion date of the calibration. All calibrations need to be successfully
performed before using the device.
NOTICE! Calibration images are shown on the GUI as viewed from the
sensor.
68
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
Recalibration is required.
Calibration performed
successfully.
ORTHOPANTOMOGRAPH
D519207, 19