KaVo ORTHOPANTOMOGRAPH OP 3D User And Installation Manual page 164

Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

9 Installation
The i4100 Back Panel Board contains a switch (SW1), which needs to be configured if the
type of installation is changed from permanent to plug connection.
NOTICE! If device has been delivered from the factory with plug connection, it
has already been configured accordingly.
To access the switch, you must detach the Back panel plate by removing the 2 indicated
screws.
Mains power configurations:
Permanent connection - Switch (SW1) shall be in F1 position.
Plug connection - Switch (SW1) shall be in F2 position and both fuses (F1+F2) must
be installed.
See chapter
164
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
Changing the fuses
on page 165 for information on how to install the fuses.