9 Installation
To tilt the collimator in UP/DOWN direction:
1. Loosen the 2 screws on the sides of the collimator assembly.
2. Adjust the tilt by turning the tilt adjustment screw. The amount of tilt can be
inspected from the scale at the end of the collimator.
• Turn the adjustment screw clockwise to tilt the collimator upwards.
• Turn the adjustment screw counter clockwise to tilt the collimator
downwards.
NOTICE! The GUI informs you how much the collimator needs to be
adjusted and in which direction. One full turn of the screw (360°) equals
about 0.7 mm.