9.6.2 System time check

The device receives the system time from the network it is connected to during start-up.
This chapter instructs how to check if the received time is correct and how to fix it if it is
not.
1. While in installation mode, select Settings menu and open Device time settings
tab.
2. Check that the Device time is correct.
3. If the date and time is incorrect, click Start to fetch new time from network. This will
instantly restart the device.
NOTICE! If the Start button is grayed out, restart the device from the power
switch.
4. Click Refresh to see the new system time.
NOTICE! If the date and time remains incorrect, ensure that the workstation
time is correct, NetTime service is running on the workstation and the
workstation firewall has port 123 open for UDP.
Once the device has the correct system time, continue to calibrate the device.