9.11.2 CEPH primary collimator adjustment
If the CEPH Tube Sweep Calibration has failed due to the collimator being too high/low or
tilted, adjustment is required.
NOTICE! This adjustment is required only in special cases. After the adjustment,
all cephalometric calibrations related to radiation must be performed again.
1. Detach the CEPH tubehead upper cover screw and pull the cover off.
NOTICE! The tubehead rear cover may fall down after the tubehead upper
cover is detached.
2. Pull the tubehead rear cover off.
3. Pull the tubehead front cover locking piece off and detach the tubehead front cover.