Ceph Primary Collimator Adjustment - KaVo ORTHOPANTOMOGRAPH OP 3D User And Installation Manual

Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

9.11.2 CEPH primary collimator adjustment

If the CEPH Tube Sweep Calibration has failed due to the collimator being too high/low or
tilted, adjustment is required.
NOTICE! This adjustment is required only in special cases. After the adjustment,
all cephalometric calibrations related to radiation must be performed again.
1. Detach the CEPH tubehead upper cover screw and pull the cover off.
NOTICE! The tubehead rear cover may fall down after the tubehead upper
cover is detached.
2. Pull the tubehead rear cover off.
3. Pull the tubehead front cover locking piece off and detach the tubehead front cover.
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
OP 3D
9 Installation
175
D519207, 19

Advertisement

Table of Contents
loading

Table of Contents