12. If the verification fails, ensure the alignment of the measurement device and redo
the test program. If the problem persists, contact service.
13. Adjust the verification test settings:
3D testing
• Select the length of the radiation pulse, 5ms (LDT/standard), 10ms (Standard
resolution) or Constant (High/ENDO resolution).
• Select the mA value for the test program using the + and - buttons.
14. Perform the 3D testing.

4.4.2 Positioning panel

1. Carriage up/down slider. The carriage can be driven up and down by sliding a finger
on the carriage up/down slider. The speed of the carriage movement depends on the
distance the finger is moved.
2. Test mode button. Disables radiation production of the device. You can also use the
GUI to activate the Test mode.
3. Patient positioning lights button. Turns the patient positioning lights on/off.
4. HOME button. Drives the rotating unit to HOME (Patient-In) and scan start positions.
5. Exposure indicator (YELLOW). Illuminates when the device is radiating.
6. Status indicator (BLUE). Illuminates when the device is in Error state. The error
details are shown on the GUI.
7. Ready indicator (GREEN). Illuminates when the device is ready for imaging.
NOTICE! If carriage up/down slider backlight is not illuminated, the carriage up/
down slider is in safe mode and cannot be used. Wait until the light comes back on
or wipe the panel clean if it is dirty.
ORTHOPANTOMOGRAPH
OP 3D
4 Device description
33