INSTALLATION REQUIREMENTS & ENVIRONMENT CONTROL:
In order to minimize interference risks, the following requirements shall apply.
Cables shielding & grounding
All interconnect cables to peripheral devices must meet the requirements given in
Technical specifications
causing radio frequency interference.
Electrostatic discharges environment & recommendations
In order to reduce electrostatic discharge interference, a charge dissipative floor should
be installed to prevent charge accumulation.
• The dissipative floor material must be connected to the system reference ground, if
applicable.
• Relative humidity must be maintained above 30 percent.
Stacked components & equipment
WARNING! The PCX-1 should not be used adjacent to or stacked with other
equipment; if adjacent or stacked use is necessary, the PCX-1 should be observed
to verify normal operation in the configuration in which it will be used.
Interference may occur in the vicinity of equipment marked with the following symbol:

10.7 X-ray tube assemblies

ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
on page 195. Use of incorrect cables may result in the device
OP 3D
10 Technical data
217