4. Remove the screw on the 2D/3D sensor cover.
5. Pull the sensor covers apart to detach them.
6. Loosen the 2 sensor height locking screws.
7. Turn the adjustment screw to adjust the height of the sensor. One turn is equivalent
of 0.7 mm.
8. Re-tighten the locking screws after the adjustment.
9. Redo the failed calibration to verify the adjustment.
10. Re-install the detached covers.