7.2.3 Calibration procedure

NOTICE! Some calibrations presented in this chapter are not available with all
device configurations. Perform the calibrations in the exact order as indicated on
the GUI.
After you have run all available calibrations successfully, take Quality Control images as
instructed in chapter
7.2.3.1 3D pixel calibration
This program calibrates the sensor for 3D imaging.
No calibration tools are required when performing this calibration.
NOTICE! This calibration produces X-rays. Protect yourself from radiation.
1. Select 3D pixel calibration from the Calibrations menu.
2. Protect yourself from radiation.
3. Press and hold the exposure button down to take the calibration image.
4. When the exposure warning stops and the program end tone is played, the program
is complete.
5. The calibration image appears on the GUI.
6. Acknowledge the calibration result by pressing the OK button.
7.2.3.2 PAN pixel calibration (3D sensor)
This program calibrates the 2D/3D sensor for Panoramic imaging.
No calibration tools are required when performing this calibration.
NOTICE! This calibration produces X-rays. Protect yourself from radiation.
1. Select PAN pixel calibration (3D sensor) from the Calibrations menu.
2. Protect yourself from radiation.
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
Quality Control
OP 3D
on page 73.