9 Installation

9.6.22 CEPH mechanical calibration

This program is used to check the correct alignment of the head support and the ear
rods.
No calibration tools are required when performing this calibration.
NOTICE! This calibration produces X-rays. Protect yourself from radiation.
1. Turn the CEPH head support to Lateral position.
2. Move the ear holders to the outmost position.
3. Select CEPH mechanical calibration from the Calibrations menu.
4. Protect yourself from radiation.
5. Press and hold the exposure button down to take the calibration image.
6. When the exposure warning stops and the program end tone is played, the program
is complete.
7. The calibration image appears on the GUI.
NOTICE! The ear rods have small metal balls at their tips. The calibration
passes when the balls are close enough to each other.
NOTICE! If the calibration fails and the GUI informs you to adjust the ear
rods or the ear rod assembly, perform the required adjustments according to
chapter
8. Acknowledge the calibration result by pressing the OK button.
NOTICE! Repeat the CEPH head support calibration if the GUI requests it.
140
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
Ear rod adjustment
on page 190.