Pan Collimator Calibration - KaVo ORTHOPANTOMOGRAPH OP 3D User And Installation Manual

Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

6. The calibration results appear on the GUI.
7. Acknowledge the calibration by pressing the OK button.
NOTICE! If the calibration fails and the GUI doesn't tell you to do any
adjustments, repeat the calibration.
NOTICE! If the calibration fails and the GUI informs you to adjust the sensor,
adjust the 2D sensor according to chapter

9.6.13 PAN collimator calibration

This program calibrates the PAN collimator.
No calibration tools are required when performing this calibration.
NOTICE! This calibration produces X-rays. Protect yourself from radiation.
1. Select PAN collimator calibration from the Calibrations menu.
2. Protect yourself from radiation.
3. Press and hold the exposure button down to take the calibration image.
4. When the exposure warning stops and the program end tone is played, the program
is complete.
ORTHOPANTOMOGRAPH
Reviewed Suuronen Esa 2019-09-19 12:21
Approved Lassinen Juuso 2019-10-02 14:37
OP 3D
2D Sensor adjustments
9 Installation
on page 182.
129
D519207, 19