You can use a well loop to join together a selection of protocol steps and to
redefine the protocol parameters for them. The steps are run in the order
specified in the Steps step tree in a loop until all the wells are measured.
You can specify the number of wells in the loop or you can select the whole
plate.
In addition, you can control the execution order of kinetic measurements,
kinetic dispense and measurement steps, and dispense and measurement
steps.
Block interval [hh:mm:ss] - Enter the duration of the well loop
execution. If no block interval is defined, the instrument performs
consequent loops with no delay in between.
Well interval [ms] - Enter the time that elapses between the starting
times of two consecutive wells. For example, to keep the time between
the measurement after dispensing and the next measurement constant
in a kinetic loop. If selected, the value must be at least 10 ms greater
than the Settle delay value. This option is not available when the
execution block type is set to Well and the well count is 1 (see below).
Thermo Scientific SkanIt
®
®
Software for Varioskan
User Manual
Protocols
Well loops
87