Thermo Scientific Varioskan User Manual page 191

Multimode microplate reader software

Advertisement

Thermo Fisher Scientific
You can define the following settings:
Reaction - Define whether the reaction is supposed to produce
increasing or decreasing signal levels. There are three options for this
setting (see Maximum rate" on page 185):
Undefined - Searches for the absolute maximum rate, calculates the
half of the maximum rate, and returns the corresponding time as a
positive number.
Increasing - Searches for the maximum increasing rate, calculates
the half of the maximum rate, and returns the corresponding time
as a positive number. If no increasing rate is found, the result is
NaN" (not a number).
Decreasing - Searches for the maximum decreasing rate, calculates
the half of the maximum rate, and returns the corresponding time
as a positive number. If no decreasing rate is found, the result is
NaN" (not a number).
Ignore first (n) readings - Ignores a defined number of points
(readings), counted from the first reading.
Ignore last (n) readings - Ignores a defined number of points (readings),
counted from the last reading.
Thermo Scientific SkanIt
Data Processing Parameters
®
®
Software for Varioskan
User Manual
Kinetic calculations
189