Multipoint Calculation - Thermo Scientific Varioskan User Manual

Multimode microplate reader software

Advertisement

Data Processing Parameters

Multipoint calculation

Multipoint calculation
152
Thermo Scientific SkanIt
Source - Select which sample is used as the B
contains that sample. The mean value of the selected sample (if replicates
do exist) will be used in the calculation.
The B
and you can define which specific sample should be used as B
example, Cal_5 or Unknown_56. The ratio or inhibition value is
calculated for all the samples belonging to a certain group.
Multipoint calculation is used to reduce multiple measurement points in
each well to just one point per well (see Using multipoint in a
measurement" on page 114). Multipoint calculation must be performed
before any other data reduction, except Basic Statistics or Blank Subtraction,
is attempted. Multipoint calculation is only available for 6- to 48-well plates.
Select Data Processing > Multipoint Calculation, click the
the Results action panel, or right-click the measurement in the Results tree.
Average - The average value of all the measurement points in each well.
Minimum - The minimum value of all the measurement points in each
well.
Maximum - The maximum value of all the measurement points in each
well.
Maximum - Minimum - The minimum value of all the measurement
points in each well is subtracted from the maximum value of all
measurement points in the same well.
Standard deviation - The standard deviation of all the measurement
points in each well.
Sum - The sum of all the measurement points in each well.
®
®
Software for Varioskan
User Manual