Priming dispensers

Thermo Fisher Scientific
If the measurement data is not within the acceptable quality control
range, you can still choose to accept and save the K-factor.
Caution Measurements made with such K-factors will cause
unreliable PLC measurement results.
7.
Give a name to the solvent and add an optional description for the
solvent. The K-factor is automatically entered with a 3-decimal
accuracy. Click OK. The pathlength correction factor is saved and
added to the Pathlength Correction K-Factors dialog that is launched
when the pathlength correction is selected in a photometric
measurement (see Using pathlength correction in a measurement"
on page 97).
You can perform calculations with the pathlength correction measurement
data, if you wish. First add a User Equation calculation step and then any
other calculation step, such as Quality Control.
Caution You must always prime into a waste vessel by removing
the dispensing head from the dispensing head positioner. Hold on
to the dispensing head while priming.
Select Execute > Prime... or click the
Thermo Scientific SkanIt
button.
®
®
Software for Varioskan
User Manual
Executing a Session
Priming dispensers
75