Thermo Scientific Varioskan User Manual page 276

Multimode microplate reader software

Advertisement

Troubleshooting Guide
Warning messages
274
Thermo Scientific SkanIt
Code
106
107
108
109
110
®
®
Software for Varioskan
User Manual
Explanation
The lamp lifetime has reached its
end.
Protocol time settings did not
allow internal runtime calibration.
The command has no effect.
The synchronization of
measurements failed. The second
measurement started later than
requested.
Dark level interpolation buffer
overflow.
Suggested action
Arrange for the replacement of the
lamp as soon as convenient. Contact
service.
The accuracy of the measurement
results may have suffered. If there is
no waiting time in the assay, then you
have to accept the possible accuracy
reduction.
This just informs that a command has
been used which has no effect in the
current measurement method.
The minimum timing depends on the
combination of measurement
parameters, plate movement
parameters, dispensing parameters
and volume, number of wavelengths
and number of measured points.
Reduce well group size, use dark level
interpolation between wells or do not
use interpolation at all.
Thermo Fisher Scientific